Abstract
This study compared two MALDI-TOF MS systems (Biotyper and VITEK MS) on clinical Streptococcus bovis group isolates (n=66). The VITEK MS gave fewer misidentifications and a higher rate of correct identifications than the Biotyper. Only the identification of S. lutetiensis by the VITEK MS was reliable. Additional optimization of the available system databases is needed.
Original language | English |
---|---|
Journal | Diagnostic Microbiology and Infectious Disease |
Volume | 88 |
Issue number | 1 |
Pages (from-to) | 23-25 |
ISSN | 0732-8893 |
DOIs | |
Publication status | Published - 1 May 2017 |