Abstract
This study compared two MALDI-TOF MS systems (Biotyper and VITEK MS) on clinical Streptococcus bovis group isolates (n=66). The VITEK MS gave fewer misidentifications and a higher rate of correct identifications than the Biotyper. Only the identification of S. lutetiensis by the VITEK MS was reliable. Additional optimization of the available system databases is needed.
Originalsprog | Engelsk |
---|---|
Tidsskrift | Diagnostic Microbiology and Infectious Disease |
Vol/bind | 88 |
Udgave nummer | 1 |
Sider (fra-til) | 23-25 |
ISSN | 0732-8893 |
DOI | |