Translational and rotational intra- and inter-fractional errors in patient and target position during a short course of frameless stereotactic body radiotherapy

Mirjana Josipovic, Gitte Fredberg Persson, Ashildur Logadottir, Bob Smulders, Gunnar Westmann, Jens Peter Bangsgaard

33 Citations (Scopus)

Abstract

BACKGROUND: Implementation of cone beam computed tomography (CBCT) in frameless stereotactic body radiotherapy (SBRT) of lung tumours enables setup correction based on tumour position. The aim of this study was to compare setup accuracy with daily soft tissue matching to bony anatomy matching and evaluate intra- and inter-fractional translational and rotational errors in patient and target positions.

MATERIAL AND METHODS: Fifteen consecutive SBRT patients were included in the study. Vacuum cushions were used for immobilisation. SBRT plans were based on midventilation phase of four-dimensional (4D)-CT or three-dimensional (3D)-CT from PET/CT. Margins of 5 mm in the transversal plane and 10 mm in the cranio-caudal (CC) direction were applied. SBRT was delivered in three fractions within a week. At each fraction, CBCT was performed before and after the treatment. Setup accuracy comparison between soft tissue matching and bony anatomy matching was evaluated on pretreatment CBCTs. From differences in pre- and post-treatment CBCTs, we evaluated the extent of translational and rotational intra-fractional changes in patient position, tumour position and tumour baseline shift. All image registration was rigid with six degrees of freedom.

RESULTS: The median 3D difference between patient position based on bony anatomy matching and soft tissue matching was 3.0 mm (0-8.3 mm). The median 3D intra-fractional change in patient position was 1.4 mm (0-12.2 mm) and 2.2 mm (0-13.2 mm) in tumour position. The median 3D intra-fractional baseline shift was 2.2 mm (0-4.7 mm). With correction of translational errors, the remaining systematic and random errors were approximately 1°.

CONCLUSION: . Soft tissue tumour matching improved precision of treatment delivery in frameless SBRT of lung tumours compared to image guidance using bone matching. The intra-fractional displacement of the target position was affected by both translational and rotational changes in tumour baseline position relative to the bony anatomy and by changes in patient position.

Original languageEnglish
JournalActa Oncologica
Volume51
Issue number5
Pages (from-to)610-617
ISSN0284-186X
DOIs
Publication statusPublished - May 2012

Keywords

  • Adult
  • Aged
  • Aged, 80 and over
  • Algorithms
  • Carcinoma, Non-Small-Cell Lung/pathology
  • Cone-Beam Computed Tomography
  • Dose Fractionation
  • Female
  • Follow-Up Studies
  • Humans
  • Immobilization
  • Lung Neoplasms/pathology
  • Male
  • Melanoma/pathology
  • Middle Aged
  • Movement
  • Multimodal Imaging
  • Patient Positioning
  • Positron-Emission Tomography
  • Prognosis
  • Radiosurgery
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy, Computer-Assisted
  • Rectal Neoplasms/pathology
  • Stomach Neoplasms/pathology
  • Survival Rate
  • Tomography, X-Ray Computed

Fingerprint

Dive into the research topics of 'Translational and rotational intra- and inter-fractional errors in patient and target position during a short course of frameless stereotactic body radiotherapy'. Together they form a unique fingerprint.

Cite this