Abstract
Lead isotope ratios in archaeological silver and copper were determined by MC-ICPMS using laser ablation and bulk dissolution without lead purification. Laser ablation results on high-lead metals and bulk solution analyses on all samples agree within error of TIMS data, suggesting that problems from isobaric interferences and/or mass bias variations due to the presence of matrix elements are insignificant. Inaccurate laser ablation analyses on low-lead copper reflect erroneous mass bias corrections from use of a non-matrix matched standard. However, in most cases, silver and copper are analysable for lead isotopes by bulk dissolution or laser ablation MC-ICPMS with simplified sample preparation.
Original language | English |
---|---|
Journal | Archaeometry |
Issue number | 48 |
Pages (from-to) | 45-56 |
ISSN | 0003-813X |
Publication status | Published - 2006 |