Optimizing clinical performance and geometrical robustness of a new electrode device for intracranial tumor electroporation

Faisal Mahmood, Julie Gehl

36 Citations (Scopus)

Abstract

Current technology has limited applicability for electroporation based treatment of deep-seated tumors, and is in general, not optimized in terms of compliance with clinically relevant parameters. Here we present a novel electrode device developed for electrotransfer of antineoplastic drugs and genes to intracranial tumors in humans, and demonstrate a method to optimize the design (i.e. geometry) of the electrode device prototype to improve both clinical performance and geometrical tolerance (robustness). We have employed a semiempirical objective function based on constraints similar to those used in radiation oncology. The results show that small geometrical changes may yield a significant improvement. For example, a 2 mm displacement of 6 electrodes yields 14% better compliance with the clinical parameters, compared to the prototype, and additionally makes the electrode device less sensitive to random geometrical deviations. The method is readily applicable to other electrode configurations.
Original languageEnglish
JournalBioelectrochemistry
Volume81
Issue number1
Pages (from-to)10-16
Number of pages7
ISSN1567-5394
DOIs
Publication statusPublished - Apr 2011

Fingerprint

Dive into the research topics of 'Optimizing clinical performance and geometrical robustness of a new electrode device for intracranial tumor electroporation'. Together they form a unique fingerprint.

Cite this