TY - JOUR
T1 - Indirect Calorimetry in Mechanically Ventilated Patients
T2 - A Prospective, Randomized, Clinical Validation of 2 Devices Against a Gold Standard
AU - Allingstrup, Matilde Jo
AU - Kondrup, Jens
AU - Perner, Anders
AU - Christensen, Poul Lunau
AU - Jensen, Tom Hartvig
AU - Henneberg, Steen Winther
PY - 2017/11
Y1 - 2017/11
N2 - Background and Aims: The 2 currently available indirect calorimeters, CCM Express Indirect Calorimeter (MedGraphics, St Paul, MN) and Quark RMR ICU Indirect Calorimeter (COSMED, Rome, Italy), have not been validated against a gold standard in mechanically ventilated patients. Our aim was to do so using a gold-standard, modified Tissot bell-spirometer method in mechanically ventilated patients who were hemodynamically, respiratory, and metabolically stable. Methods: We studied 30 patients undergoing general anesthesia and major gynecological surgery. We measured oxygen consumption ((Formula presented.) O2) and resting energy expenditure (REE) in a randomized, sequential, crossover design with double determination of each device. Results: Compared with the modified Tissot bell-spirometer, the CCM Express Indirect Calorimeter demonstrated a mean Δ-REE of +361 kcal/d, corresponding to a 31% overestimation of energy requirements. Bland-Altman analysis for REE showed a mean (SD) bias of 384 (124) with limits of agreement 142–627 kcal/d. QUARK RMR ICU demonstrated a mean Δ-REE of 81 kcal/d, corresponding to a 7% overestimation of energy requirements. Bland-Altman plot analysis showed a mean (SD) bias of 77 (167) with limits of agreement −249 to 404 kcal/d. Conclusions: The QUARK RMR ICU Indirect Calorimeter compared better with the gold standard for values of (Formula presented.) O2 and REE than did the CCM Express Indirect Calorimeter in mechanically ventilated patients who were circulatory and respiratory stable. Both indirect calorimeters had low precision.
AB - Background and Aims: The 2 currently available indirect calorimeters, CCM Express Indirect Calorimeter (MedGraphics, St Paul, MN) and Quark RMR ICU Indirect Calorimeter (COSMED, Rome, Italy), have not been validated against a gold standard in mechanically ventilated patients. Our aim was to do so using a gold-standard, modified Tissot bell-spirometer method in mechanically ventilated patients who were hemodynamically, respiratory, and metabolically stable. Methods: We studied 30 patients undergoing general anesthesia and major gynecological surgery. We measured oxygen consumption ((Formula presented.) O2) and resting energy expenditure (REE) in a randomized, sequential, crossover design with double determination of each device. Results: Compared with the modified Tissot bell-spirometer, the CCM Express Indirect Calorimeter demonstrated a mean Δ-REE of +361 kcal/d, corresponding to a 31% overestimation of energy requirements. Bland-Altman analysis for REE showed a mean (SD) bias of 384 (124) with limits of agreement 142–627 kcal/d. QUARK RMR ICU demonstrated a mean Δ-REE of 81 kcal/d, corresponding to a 7% overestimation of energy requirements. Bland-Altman plot analysis showed a mean (SD) bias of 77 (167) with limits of agreement −249 to 404 kcal/d. Conclusions: The QUARK RMR ICU Indirect Calorimeter compared better with the gold standard for values of (Formula presented.) O2 and REE than did the CCM Express Indirect Calorimeter in mechanically ventilated patients who were circulatory and respiratory stable. Both indirect calorimeters had low precision.
KW - indirect calorimetry
KW - intensive care
KW - mechanical ventilation
KW - oxygen consumption
KW - resting energy expenditure (REE)
KW - validation
U2 - 10.1177/0148607116662000
DO - 10.1177/0148607116662000
M3 - Journal article
C2 - 27488830
AN - SCOPUS:85035023590
SN - 0148-6071
VL - 41
SP - 1272
EP - 1277
JO - Journal of Parenteral and Enteral Nutrition
JF - Journal of Parenteral and Enteral Nutrition
IS - 8
ER -