Electromagnetic source imaging in presurgical workup of patients with epilepsy: A prospective study

Lene Duez, Hatice Tankisi, Peter Orm Hansen, Per Sidenius, Anne Sabers, Lars H Pinborg, Martin Fabricius, György Rásonyi, Guido Rubboli, Birthe Pedersen, Anne-Mette Leffers, Peter Uldall, Bo Jespersen, Jannick Brennum, Otto Mølby Henriksen, Anders Fuglsang-Frederiksen, Sándor Beniczky

24 Citations (Scopus)
73 Downloads (Pure)

Abstract

OBJECTIVE: To determine the diagnostic accuracy and clinical utility of electromagnetic source imaging (EMSI) in presurgical evaluation of patients with epilepsy.

METHODS: We prospectively recorded magnetoencephalography (MEG) simultaneously with EEG and performed EMSI, comprising electric source imaging, magnetic source imaging, and analysis of combined MEG-EEG datasets, using 2 different software packages. As reference standard for irritative zone (IZ) and seizure onset zone (SOZ), we used intracranial recordings and for localization accuracy, outcome 1 year after operation.

RESULTS: We included 141 consecutive patients. EMSI showed localized epileptiform discharges in 94 patients (67%). Most of the epileptiform discharge clusters (72%) were identified by both modalities, 15% only by EEG, and 14% only by MEG. Agreement was substantial between inverse solutions and moderate between software packages. EMSI provided new information that changed the management plan in 34% of the patients, and these changes were useful in 80%. Depending on the method, EMSI had a concordance of 53% to 89% with IZ and 35% to 73% with SOZ. Localization accuracy of EMSI was between 44% and 57%, which was not significantly different from MRI (49%-76%) and PET (54%-85%). Combined EMSI achieved significantly higher odds ratio compared to electric source imaging and magnetic source imaging.

CONCLUSION: EMSI has accuracy similar to established imaging methods and provides clinically useful, new information in 34% of the patients.

CLASSIFICATION OF EVIDENCE: This study provides Class IV evidence that EMSI had a concordance of 53%-89% and 35%-73% (depending on analysis) for the localization of epileptic focus as compared with intracranial recordings-IZ and SOZ, respectively.

Original languageEnglish
JournalNeurology
Volume92
Issue number6
Pages (from-to)e576-e586
ISSN0028-3878
DOIs
Publication statusPublished - 5 Feb 2019

Fingerprint

Dive into the research topics of 'Electromagnetic source imaging in presurgical workup of patients with epilepsy: A prospective study'. Together they form a unique fingerprint.

Cite this