Comparing twelve-lead electrocardiography with close-to-heart patch based electrocardiography

Ingeborg H Hansen, Karsten Hoppe, Anna Gjerde, Joergen K Kanters, Helge B D Sorensen

9 Citations (Scopus)

Abstract

Electrocardiographic (ECG) recording using adhesive patch-type ECG monitors (PEMs) has several advantages over conventional ECG recorders. However, due to the unconventional electrode locations used in PEM systems, the morphology of the acquired ECG signals may differ from conventional ECG leads used in the clinic impeding clinical interpretation. In this study, recordings from an ePatch® lead system involving three torso sites are compared with concurrently recorded standard 12-lead ECG. Pearson's correlation coefficients (CC) of -0.90 and 0.91 is found between two of the PEM signals and the standard 12-lead ECG signals aVR and V2, respectively. Deriving the 12-lead ECG from the PEM leads through linear transforms on a subject-specific basis yield CC values ranging from 0.78 to 0.96 between measured and derived leads. The corresponding CC values for the PEM ECG leads range from 0.88 to 0.95. It is found that the PEM lead system captures "residual" information not contained in the standard 12-lead ECG and i.a. a negative deflection after the T-wave is discovered in the PEM signals.

Original languageEnglish
JournalI E E E Engineering in Medicine and Biology Society. Conference Proceedings
Volume2015
Pages (from-to)330-333
Number of pages4
ISSN2375-7477
DOIs
Publication statusPublished - 4 Nov 2015