Automated method for the direct analysis of 8-oxo-guanosine and 8-oxo-2'-deoxyguanosine in human urine using ultraperformance liquid chromatography and tandem mass spectrometry

Trine Henriksen, Peter R Hillestrøm, Henrik E Poulsen, Allan Weimann, Trine Henriksen, Peter René Hillestrøm, Henrik E Poulsen, Allan Weimann

    79 Citations (Scopus)

    Abstract

    The potential use of oxidative stress-induced DNA and RNA damage products as biomarkers is an important aspect of biomedical research. There is a need for assays with high specificity and sensitivity that also can be used in molecular epidemiology studies with a large number of subjects. In addition there is a need for assays that can measure more than one product from DNA oxidation. We present a sensitive, precise, and accurate method for quantitative analysis of the oxidized nucleosides 8-oxoGuo and 8-oxodG in human urine. The assay is based on automated sample handling using a BIOMEK 3000 Workstation, and UPLC-ESI(+)-MS/MS analysis. High specificity is evidenced by the use of qualifier ions for both analytes. The quantification limit in urine samples is 1 nM for both analytes. Accuracy and precision were documented, showing average recoveries of 106.2% (8-oxoGuo) and 106.9% (8-oxodG), and overall precision (within-day and between-days) of 6.1 and 4.4%, respectively.
    Original languageEnglish
    JournalFree Radical Biology & Medicine
    Volume47
    Issue number5
    Pages (from-to)629-35
    Number of pages7
    ISSN0891-5849
    DOIs
    Publication statusPublished - 1 Sept 2009

    Fingerprint

    Dive into the research topics of 'Automated method for the direct analysis of 8-oxo-guanosine and 8-oxo-2'-deoxyguanosine in human urine using ultraperformance liquid chromatography and tandem mass spectrometry'. Together they form a unique fingerprint.

    Cite this