Assessment of left ventricular size and function during cardiac surgery: An intraoperative evaluation of six two-dimensional echocardiographic methods with real time three-dimensional echocardiography as a reference

M. Grossgasteiger, B. Graser, H. Rauch, J. Motsch, M.D. Hien, Matthias Gondan, C. Rosendal

13 Citations (Scopus)

Abstract

Transesophageal echocardiography is recommended to monitor left ventricular (LV) size and function in various operations. Generally, two-dimensional (2D) methods are applied intraoperatively. The aim of this study was to compare the accuracy and feasibility of 6 commonly used 2D methods to assess LV function during surgery. LV function in 120 consecutive patients was evaluated. Real time three-dimensional transesophageal echocardiograpy (3DTEE) served as reference. End-diastolic and end-systolic volumes and ejection fraction (EF) were analyzed with Simpson's method of discs (monoplane [MP] and biplane [BP]), eyeball method, Teichholz' method, and speckle tracking (ST) methods. Furthermore, fractional area change (FAC) and peak systolic pressure rise (dP/dt) were determined. Each 2D method was evaluated regarding correlation and agreement with 3DE, intra- and interobserver variability and the time required for evaluation. Simpson BP showed the strongest correlation and best agreement with 3DE for EF (limits of agreement 3.7 ± 11.6%) and volumes. Simpson MP showed similar agreement with 3DE compared to ST (2.8 ± 14.5% vs. 2.0 ± 15.3% and 3.8 ± 14.4% vs. 1.9 ± 15.6%, respectively). Both the eyeball method and Teichholz' method showed wide limits of agreement (-1.5 ± 18.2% and 5.2 ± 22.1%, respectively). dP/dt did not correlate with 3DE. FAC and ST FAC showed similar agreement. Application of 3DE (429 ± 108 seconds) took the longest time, and the eyeball method took the shortest time (8 ± 5 seconds) for analysis. Simpson BP is the most accurate intraoperative 2D method to evaluate LV function, followed by long-axis MP evaluations. Short-axis views were less accurate but may be suited for monitoring. We do not recommend using dP/dt.
Original languageEnglish
JournalEchocardiography: A Journal of Cardiovascular Ultrasound and Allied Techniques
Volume30
Issue number6
Pages (from-to)672-681
Number of pages10
ISSN0742-2822
DOIs
Publication statusPublished - 1 Jul 2013
Externally publishedYes

Fingerprint

Dive into the research topics of 'Assessment of left ventricular size and function during cardiac surgery: An intraoperative evaluation of six two-dimensional echocardiographic methods with real time three-dimensional echocardiography as a reference'. Together they form a unique fingerprint.

Cite this