Water vapor inhibits hydrogen sulfide detection in pulsed fluorescence sulfur monitors

Anders Brostrøm Bluhme, Jonas L. Ingemar, Carl Meusinger, Matthew Stanley Johnson

3 Citationer (Scopus)
60 Downloads (Pure)

Abstract

The Thermo Scientific 450 Hydrogen Sulfide- Sulfur Dioxide Analyzer measures both hydrogen sulfide (H2S) and sulfur dioxide (SO2). Sulfur dioxide is measured by pulsed fluorescence, while H2S is converted to SO2 with a molybdenum catalyst prior to detection. The 450 is widely used to measure ambient concentrations, e.g., for emissions monitoring and pollution control. An air stream with a constant H2S concentration was generated and the output of the analyzer recorded as a function of relative humidity (RH). The analyzer underreported H2S as soon as the relative humidity was increased. The fraction of undetected H2S increased from 8.3 at 5.3% RH (294 K) to over 34% at RH>80 %. Hydrogen sulfide mole fractions of 573, 1142, and 5145 ppb were tested. The findings indicate that previous results obtained with instruments using similar catalysts should be re-evaluated to correct for interference from water vapor. It is suspected that water decreases the efficiency of the converter unit and thereby reduces the measured H2S concentration.