Abstract
Traditionally, segmentation between multi-materials in CT is only available for cases, where material densities are not close to each other. A novel method called GBI offers a new possibility to overcome this problem, and was evaluated with respect to its metrological performance by comparisons to traceable measurements. The measurement results show that further development related to stability issues on the used CT is needed to create a metrological tool using GBI.
Originalsprog | Engelsk |
---|---|
Titel | Proceedings of the 14th euspen International Conference : 02/06/14 → 06/06/14, Dubrovnik, Croatia |
Udgivelsessted | Dubrovnik, Croatia |
Publikationsdato | 2 jun. 2014 |
Udgave | euspen |
Sider | 157-160 |
Status | Udgivet - 2 jun. 2014 |