Test of Investigator ESSPLEX SE QS with quality sensors

Abstract

The Investigator ESSplex SE QS (Qiagen) amplifies sixteen STRs and Amelogenin. The reaction mix is supplemented with two quality sensors that are sensitive to inhibitors in the amplified samples. The quality sensors consist of a short fragment and a long fragment. Amplification of the long fragment is very sensitive to inhibitors, while the short fragment is very resistant to inhibitors. The kit sensitivity was tested with serial dilutions of high quality DNA. Full profiles were obtained down to 125 pg of DNA. The behaviour of the quality sensors was investigated by adding Hematin or Indigo Carmine to high quality DNA samples. The quality sensor peak height ratio increased and the peak heights of the longer fragments of the profiles decreased. The decreasing peak heights of the longer fragments are classical signs of DNA degradation. With artificially degraded DNA, the quality sensor peak height ratio was not affected. Investigations of 16 DNA extracts from stain samples from crime cases indicated that six of the samples were contaminated with PCR inhibitors and 10 of the samples contained degraded DNA.

OriginalsprogEngelsk
TidsskriftForensic Science International: Genetics. Supplement Series
Vol/bind5
Sider (fra-til)e490–e491
Antal sider2
ISSN1875-1768
DOI
StatusUdgivet - 1 dec. 2015

Fingeraftryk

Dyk ned i forskningsemnerne om 'Test of Investigator ESSPLEX SE QS with quality sensors'. Sammen danner de et unikt fingeraftryk.