Abstract
BACKGROUND: Periacetabular osteotomy (PAO) is the treatment of choice for younger patients with developmental hip dysplasia. The procedure aims to normalize the joint configuration, reduce the peak-pressure, and delay the development of osteoarthritis. The procedure is technically demanding and no previous study has validated the use of computer navigation with a minimally invasive transsartorial approach.
METHODS: Computer-assisted PAO was performed on ten patients. Patients underwent pre- and postoperative computed tomography (CT) scanning with a standardized protocol. Preoperative preparation consisted of outlining the lunate surface and segmenting the pelvis and femur from CT data. The Biomechanical Guidance System was used intra-operatively to automatically calculate diagnostic angles and peak-pressure measurements. Manual diagnostic angle measurements were performed based on pre- and postoperative CT. Differences in angle measurements were investigated with summary statistics, intraclass correlation coefficient, and Bland-Altman plots. The percentage postoperative change in peak-pressure was calculated.
RESULTS: Intra-operative reported angle measurements show a good agreement with manual angle measurements with intraclass correlation coefficient between 0.94 and 0.98. Computer navigation reported angle measurements were significantly higher for the posterior sector angle ([Formula: see text], [Formula: see text]) and the acetabular anteversion angle ([Formula: see text], [Formula: see text]). No significant difference was found for the center-edge ([Formula: see text]), acetabular index ([Formula: see text]), and anterior sector angle ([Formula: see text]). Peak-pressure after PAO decreased by a mean of 13% and was significantly different ([Formula: see text]).
CONCLUSIONS: We found that computer navigation can reliably be used with a minimally invasive transsartorial approach PAO. Angle measurements generally agree with manual measurements and peak-pressure was shown to decrease postoperatively. With further development, the system will become a valuable tool in the operating room for both experienced and less experienced surgeons performing PAO. Further studies with a larger cohort and follow-up will allow us to investigate the association with peak-pressure and postoperative outcome and pave the way to clinical introduction.
Originalsprog | Engelsk |
---|---|
Tidsskrift | International Journal of Computer Assisted Radiology and Surgery |
Vol/bind | 13 |
Udgave nummer | 12 |
Sider (fra-til) | 2021-2028 |
Antal sider | 8 |
ISSN | 1861-6410 |
DOI | |
Status | Udgivet - 2018 |