Methods for motion correction evaluation using 18F-FDG human brain scans on a high-resolution PET scanner

Sune H Keller, Merence Sibomana, Oline V Olesen, Claus Svarer, Søren Holm, Flemming L Andersen, Liselotte Højgaard

    27 Citationer (Scopus)

    Abstract

    Many authors have reported the importance of motion correction (MC) for PET. Patient motion during scanning disturbs kinetic analysis and degrades resolution. In addition, using misaligned transmission for attenuation and scatter correction may produce regional quantification bias in the reconstructed emission images. The purpose of this work was the development of quality control (QC) methods for MC procedures based on external motion tracking (EMT) for human scanning using an optical motion tracking system. Methods: Two scans with minor motion and 5 with major motion (as reported by the optical motion tracking system) were selected from 18F-FDG scans acquired on a PET scanner. The motion was measured as the maximum displacement of the markers attached to the subject's head and was considered to be major if larger than 4 mm and minor if less than 2 mm. After allowing a 40- to 60-min uptake time after tracer injection, we acquired a 6-min transmission scan, followed by a 40-min emission list-mode scan. Each emission list-mode dataset was divided into 8 frames of 5 min. The reconstructed time-framed images were aligned to a selected reference frame using either EMT or the AIR (automated image registration) software. The following 3 QC methods were used to evaluate the EMT and AIR MC: a method using the ratio between 2 regions of interest with gray matter voxels (GM) and white matter voxels (WM), called GM/WM; mutual information; and cross correlation. Results: The results of the 3 QC methods were in agreement with one another and with a visual subjective inspection of the image data. Before MC, the QC method measures varied significantly in scans with major motion and displayed limited variations on scans with minor motion. The variation was significantly reduced and measures improved after MC with AIR, whereas EMT MC performed less well. Conclusion: The 3 presented QC methods produced similar results and are useful for evaluating tracer-independent external-tracking motion-correction methods for human brain scans.

    OriginalsprogEngelsk
    TidsskriftJournal of Nuclear Medicine
    Vol/bind53
    Udgave nummer3
    Sider (fra-til)495-504
    Antal sider10
    ISSN0161-5505
    DOI
    StatusUdgivet - 1 mar. 2012

    Fingeraftryk

    Dyk ned i forskningsemnerne om 'Methods for motion correction evaluation using 18F-FDG human brain scans on a high-resolution PET scanner'. Sammen danner de et unikt fingeraftryk.

    Citationsformater