Intra-Operative Vector Flow Imaging Using Ultrasound of the Ascending Aorta among 40 Patients with Normal, Stenotic and Replaced Aortic Valves

Kristoffer Lindskov Hansen, Hasse Møller-Sørensen, Jesper Kjaergaard, Maiken Brit Jensen, Jens Teglgaard Lund, Mads Møller Pedersen, Theis Lange, Jørgen Arendt Jensen, Michael Bachmann Nielsen

27 Citationer (Scopus)

Abstract

Stenosis of the aortic valve gives rise to more complex blood flows with increased velocities. The angle-independent vector flow ultrasound technique transverse oscillation was employed intra-operatively on the ascending aorta of (I) 20 patients with a healthy aortic valve and 20 patients with aortic stenosis before (IIa) and after (IIb) valve replacement. The results indicate that aortic stenosis increased flow complexity (p < 0.0001), induced systolic backflow (p < 0.003) and reduced systolic jet width (p < 0.0001). After valve replacement, the systolic backflow and jet width were normalized (p < 0.52 and p < 0.22), but flow complexity was not (p < 0.0001). Flow complexity (p < 0.0001), systolic jet width (p < 0.0001) and systolic backflow (p < 0.001) were associated with peak systolic velocity. The study found that aortic stenosis changes blood flow in the ascending aorta and valve replacement corrects some of these changes. Transverse oscillation may be useful for assessment of aortic stenosis and optimization of valve surgery.

OriginalsprogEngelsk
TidsskriftUltrasound in Medicine & Biology
Vol/bind42
Udgave nummer10
Sider (fra-til)2414-2422
Antal sider9
ISSN0301-5629
DOI
StatusUdgivet - okt. 2016

Fingeraftryk

Dyk ned i forskningsemnerne om 'Intra-Operative Vector Flow Imaging Using Ultrasound of the Ascending Aorta among 40 Patients with Normal, Stenotic and Replaced Aortic Valves'. Sammen danner de et unikt fingeraftryk.

Citationsformater