Abstract
Measurements of the radiance distribution and fluence rate within turbid samples with fiber-optic radiance microprobes contain a large variable instrumental error caused by the nonuniform directional sensitivity of the microprobes. A general theory of three-dimensional radiance measurements is presented that provides correction for this error by using the independently obtained function of the angular sensitivity of the microprobes.
Originalsprog | Engelsk |
---|---|
Tidsskrift | Applied Optics |
Vol/bind | 36 |
Udgave nummer | 25 |
Sider (fra-til) | 6520-6528 |
Antal sider | 9 |
ISSN | 1559-128X |
DOI | |