Design and test of a flexible electrochemical setup for measurements in aqueous electrolyte solutions at elevated temperature and pressure

Gustav Karl Henrik Wiberg, Michael Fleige, Matthias Arenz

13 Citationer (Scopus)

Abstract

We present a detailed description of the construction and testing of an electrochemical cell allowing measurements at elevated temperature and pressure. The cell consists of a stainless steel pressure vessel containing the electrochemical glass cell exhibiting a three electrode configuration. The design of the working electrode is inspired by conventional rotating disk electrode setups. As demonstrated, the setup can be used to investigate temperature dependent electrochemical processes on polycrystalline platinum and also high surface area type electrocatalysts.

OriginalsprogEngelsk
Artikelnummer085105
TidsskriftReview of Scientific Instruments
Vol/bind85
Antal sider5
ISSN0034-6748
DOI
StatusUdgivet - aug. 2014