Abstract
Background
Compression anastomotic ring-locking procedure (CARP) is a novel procedure for creating colonic anastomoses. The surgical procedure allows perioperative quantification of the compression pressure between the intestinal ends within the anastomosis and postoperative monitoring of the anastomotic integrity. We have recently shown that CARP is a safe and effective method for colonic anastomoses in pigs, and the purpose of the present study was to evaluate CARP for colonic anastomoses in humans.
Materials and methods
This is a prospective study on 25 patients undergoing elective left-sided colonic resection. Time for evacuation of the anastomotic rings, perioperative compression pressure, and adverse effects were recorded. Postoperative blood samples were collected daily, and flexible sigmoidoscopy was performed 8–12 weeks after surgery to examine the anastomoses.
Results
Fourteen out of 25 patients underwent CARP. CARP was not used in 11 patients due to advanced tumor disease (two cases) and size restrictions (nine cases). No case of anastomotic leakage, bowel obstruction, or stenosis formation was observed. No device-related perioperative adverse events were noted. The surgical device evacuated spontaneously in all patients by the natural route after a median of 10 days. Perioperative compression pressure ranged between 85 and 280 mBar (median 130 mBar). Flexible sigmoidoscopy revealed smooth anastomoses without signs of pathological inflammation or stenosis in all cases.
Conclusion
Our results indicate that the novel suture-less CARP is a safe and effective method for creating colonic anastomoses. Further studies are warranted in larger patient populations to compare CARP head-on-head with stapled and/or hand-sewn colonic anastomoses.
Compression anastomotic ring-locking procedure (CARP) is a novel procedure for creating colonic anastomoses. The surgical procedure allows perioperative quantification of the compression pressure between the intestinal ends within the anastomosis and postoperative monitoring of the anastomotic integrity. We have recently shown that CARP is a safe and effective method for colonic anastomoses in pigs, and the purpose of the present study was to evaluate CARP for colonic anastomoses in humans.
Materials and methods
This is a prospective study on 25 patients undergoing elective left-sided colonic resection. Time for evacuation of the anastomotic rings, perioperative compression pressure, and adverse effects were recorded. Postoperative blood samples were collected daily, and flexible sigmoidoscopy was performed 8–12 weeks after surgery to examine the anastomoses.
Results
Fourteen out of 25 patients underwent CARP. CARP was not used in 11 patients due to advanced tumor disease (two cases) and size restrictions (nine cases). No case of anastomotic leakage, bowel obstruction, or stenosis formation was observed. No device-related perioperative adverse events were noted. The surgical device evacuated spontaneously in all patients by the natural route after a median of 10 days. Perioperative compression pressure ranged between 85 and 280 mBar (median 130 mBar). Flexible sigmoidoscopy revealed smooth anastomoses without signs of pathological inflammation or stenosis in all cases.
Conclusion
Our results indicate that the novel suture-less CARP is a safe and effective method for creating colonic anastomoses. Further studies are warranted in larger patient populations to compare CARP head-on-head with stapled and/or hand-sewn colonic anastomoses.
Originalsprog | Engelsk |
---|---|
Tidsskrift | International Journal of Colorectal Disease |
Vol/bind | 30 |
Udgave nummer | 7 |
Sider (fra-til) | 969-75 |
Antal sider | 7 |
ISSN | 0179-1958 |
DOI | |
Status | Udgivet - 19 jul. 2015 |