Abstract
A direct comparison between two complete intensity datasets, collected on the same sample loaded in two identical diamond-anvil pressure cells equipped, respectively, with beryllium and diamond backing plates was performed. The results clearly demonstrate that the use of diamond-backing plates significantly improves the quality of crystal structure data. There is a decrease in the internal R factor for averaging, structure refinement agreement factors, and in the errors and uncertainties of the atomic coordinates, atomic displacement parameters, and individual bond lengths.
Originalsprog | Engelsk |
---|---|
Artikelnummer | 055111 |
Tidsskrift | Review of Scientific Instruments |
Vol/bind | 82 |
Antal sider | 5 |
ISSN | 0034-6748 |
DOI | |