Central and peripheral blood flow during exercise with a continuous-flow left ventricular assist device: constant versus increasing pump speed: a pilot study

Patrice Brassard, Annette S Jensen, Nikolai Nordsborg, Finn Gustafsson, Jacob E Møller, Christian Hassager, Søren Boesgaard, Peter Bo Hansen, Peter Skov Olsen, Kåre Sander, Niels H Secher, Per Lav Madsen

    66 Citationer (Scopus)

    Abstract

    Background: End-stage heart failure is associated with impaired cardiac output (CO) and organ blood flow. We determined whether CO and peripheral perfusion are maintained during exercise in patients with an axial-flow left ventricular assist device (LVAD) and whether an increase in LVAD pump speed with work rate would increase organ blood flow. Methods and Results: Invasively determined CO and leg blood flow and Doppler-determined cerebral perfusion were measured during 2 incremental cycle exercise tests on the same day in 8 patients provided with a HeartMate II LVAD. In random order, patients exercised both with a constant (≈9775 rpm) and with an increasing pump speed (+400 rpm per exercise stage). At 60 W, the elevation in CO was more pronounced with increased pump speed (8.7±0.6 versus 8.1 ± 1.1 L·min -1; mean±SD; P=0.05), but at maximal exercise increases in CO (from 7.0±0.9 to 13.6±2.5 L · min -1; P<0.01) and leg blood flow [0.7 (0.5 to 0.8) to 4.4 (3.9 to 4.8) L · min -1 per leg; median (range); P<0.001] were similar with both pumping modes. Normally, middle cerebral artery mean flow velocity increases from ≈50 to ≈65 cm · s -1 during exercise, but in LVAD patients with a constant pump speed it was low at rest (39±14 cm · s -1) and remained unchanged during exercise, whereas in patients with increasing pump speed, it increased by 5.2±2.8 cm · s -1 at 60 W (P<0.01). Conclusions: With maximal exercise, the axial-flow LVAD supports near-normal increments in cardiac output and leg perfusion, but cerebral perfusion is poor. Increased pump speed augments cerebral perfusion during exercise. (Circ

    OriginalsprogEngelsk
    TidsskriftCirculation. Heart Failure
    Vol/bind4
    Udgave nummer5
    Sider (fra-til)554-560
    Antal sider7
    ISSN1941-3289
    DOI
    StatusUdgivet - sep. 2011

    Fingeraftryk

    Dyk ned i forskningsemnerne om 'Central and peripheral blood flow during exercise with a continuous-flow left ventricular assist device: constant versus increasing pump speed: a pilot study'. Sammen danner de et unikt fingeraftryk.

    Citationsformater