Abstract
The incidence of ventricular tachycardia (VT) or ventricular fibrillation (VF) in patients supported with a continuous-flow left ventricular assist device (LVAD) has not been investigated in detail. In 23 consecutive recipients of a HeartMate II, we analyzed the incidence of VT/VF during a total of 266 months of follow-up. Sustained VT or VF occurred in 52% of the patients, with the majority of arrhythmias occurring in the first 4 weeks after LVAD implantation. VT/VF requiring implantable cardioverter-defibrillator (ICD) shock or external defibrillation occurred in 8 patients and significant hemodynamic instability ensued in 3 patients. There were no clear predictors of VT/VF, and it is argued that prophylactic ICD implantation should be considered in patients supported with a continuous-flow LVAD.